کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
876640 | 910855 | 2011 | 8 صفحه PDF | دانلود رایگان |

In this paper a self-opening intrafascicular neural interface (SELINE) has been modeled using both a theoretical approach and a Finite Element (FE) analysis. This innovative self opening interface has several potential advantages such as: higher selectivity due to its three-dimensional structure and efficient anchorage system. Mechanical, structural and micro-technological issues have been considered to obtain an effective design of the electrode, as a feasibility study of the self-opening approach. A simple framework has been provided to model the insertion and partial retraction into peripheral nerves, resulting in the opening of wings.This integrated approach results in a rational procedure to optimize kinematics, geometry, and structural properties of peripheral interfaces. The design and feasibility study carried out in this work can potentially assure a correct behavior and dimensioning of the neural interface: in this way anomalous breakage should be avoided while mechanical and geometrical biocompatibility should increase.
Journal: Medical Engineering & Physics - Volume 33, Issue 10, December 2011, Pages 1254–1261