کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
876661 910856 2009 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Biomechanical comparison between lumbar disc arthroplasty and fusion
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Biomechanical comparison between lumbar disc arthroplasty and fusion
چکیده انگلیسی

The artificial disc is a mobile implant for degenerative disc replacement that attempts to lessen the degeneration of the adjacent elements. However, inconsistent biomechanical results for the neighboring elements have been reported in a number of studies. The present study used finite element (FE) analysis to explore the biomechanical differences at the surgical and both adjacent levels following artificial disc replacement and interbody fusion procedures.First, a three-dimensional FE model of a five-level lumbar spine was established by the commercially available medical imaging software Amira 3.1.1, and FE software ANSYS 9.0. After validating the five-level intact (INT) model with previous in vitro studies, the L3/L4 level of the INT model was modified to either insert an artificial disc (ProDisc II; ADR) or incorporate bilateral posterior lumbar interbody fusion (PLIF) cages with a pedicle screw fixation system. All models were constrained at the bottom of the L5 vertebra and subjected to 150 N preload and 10 N m moments under four physiological motions.The ADR model demonstrated higher range of motion (ROM), annulus stress, and facet contact pressure at the surgical level compared to the non-modified INT model. At both adjacent levels, ROM and annulus stress were similar to that of the INT model and varied less than 7%. In addition, the greatest displacement of posterior annulus occurred at the superior-lateral region. Conversely, the PLIF model showed less ROM, less annulus stress, and no facet contact pressure at the surgical level compared to the INT model. The adjacent levels had obviously high ROM, annulus stress, and facet contact pressure, especially at the adjacent L2/3 level.In conclusion, the artificial disc replacement revealed no adjacent-level instability. However, instability was found at the surgical level, which might accelerate degeneration at the highly stressed annulus and facet joint. In contrast to disc replacement results, the posterior interbody fusion procedure revealed possibly accelerative degeneration of the annulus and facet joint at both adjacent levels.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Engineering & Physics - Volume 31, Issue 2, March 2009, Pages 244–253
نویسندگان
, , , , ,