کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
876908 910871 2009 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Battery powered neuromuscular stimulator circuit for use during simultaneous recording of myoelectric signals
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Battery powered neuromuscular stimulator circuit for use during simultaneous recording of myoelectric signals
چکیده انگلیسی

Surface Functional Electrical Stimulation (FES) requires high stimulation voltages. A step-up transformer in the output stage of the stimulation circuit is often used. In the present technical paper a voltage controlled current source (VCCS) is presented as an alternative to the transformer coupling. Two (master–slave) coupled transconductance amplifiers (TAs)—in series with pre-charged capacitors—are used to drive the output current. After each stimulation pulse the capacitors are recharged to a high voltage by a switch mode power supply (SMPS). A multiplexer in the output stage is used to provide biphasic output. Output rise-time (10–90%) was less than 2 μs at 100 mA output. Biphasic charge balanced stimulation current can be produced with a net current to ground of less than 20 nA, thus virtually separated from ground. The circuit permits recording of the volitional myoelectric signal from the stimulated muscle. It is part of a portable myoelectrically controlled FES system powered by 2 AA batteries and currently used in clinical trials.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Engineering & Physics - Volume 31, Issue 8, October 2009, Pages 1032–1037
نویسندگان
, ,