Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion?
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion?
چکیده انگلیسی
Routine VF induction and documentation of effective defibrillation still remains a reasonable part of ICD placement because an inadequate safety margin may occur in >6% of patients. The incidence of patients who were inappropriate for testing based on well-defined clinical conditions is small (<5%) in this unselected large series. Although some clinical features may predict the need for system modification, additional studies are needed to better define “acceptable efficacy” of ICDs in preventing sudden death prior to altering these standards in selected patients.