With LRMS, the ammoniated adduct ion used as a precursor ion showed an advantage for selectivity through confirmatory transitions, without affecting signal-to-noise ratios, and hence limits of detection (LODs). As also reported by some studies in the literature, methanol-based mobile phase gave better selectivity and sensitivity for the detection of P-CTXs. While the LOD for P-CTX1B and P-CTX3C met the EFSA recommendation level when using LRMS, the findings suggested careful evaluation of instrumental parameters for determination of CTXs. LODs were significantly higher for HRMS, which currently results in the need for a significantly higher sample intake. Nevertheless, HRMS allowed for the identification of artefacts and may allow for improved confirmation of the identity of P-CTXs analogues. Consequently, LRMS and HRMS are considered complementary to ensure adequate quantitation and identification of P-CTXs.