کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10549568 965624 2005 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Harmonisation of coupled calibration curves to reduce correlated effects in the analysis of natural gas by gas chromatography
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Harmonisation of coupled calibration curves to reduce correlated effects in the analysis of natural gas by gas chromatography
چکیده انگلیسی
Quantitative analysis of natural gas depends on the calibration of a gas chromatograph with certified gas mixtures and the determination of a response relationship for each species by regression analysis. The uncertainty in this calibration is dominated by variations in the amount of the sample used for each analysis that are strongly correlated for all species measured in the same run. The “harmonisation” method described here minimises the influence of these correlations on the calculated calibration curves and leads to a reduction in the root-mean-square residual deviations from the fitted curve of a factor between 2 and 5. Consequently, it removes the requirement for each run in the calibration procedure to be carried out under the same external conditions, and opens the possibility that new data, measured under different environmental or instrumental conditions, can be appended to an existing calibration database.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1062, Issue 2, 14 January 2005, Pages 239-245
نویسندگان
, , , ,