کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10554017 967949 2005 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Automated system for release studies of salicylic acid based on a SIA method
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Automated system for release studies of salicylic acid based on a SIA method
چکیده انگلیسی
The aim of this work was to describe a fully automated system for the in vitro release testing of semisolid dosage forms based on SIA technique. The system was tested for monitoring release profiles of different ointments containing 3% of salicylic acid (Belosalic, Diprosalic, Triamcinolone S). The native fluorescence of salicylic acid was used for fluorimetric detection. Phosphate buffer pH 7.4 was the receptor medium; samples were taken at 10 min intervals during 6 h of the release test; and each test was followed by calibration with five standard solutions. The linear calibration range was 0.05-10 μg ml−1 (r = 0.9996, six standards); the maximal SIA sample throughput for this system was 120 h−1, sample volume being 50 μl and flow rate 50 μl s−1. The detection limit for salicylic acid was 0.01 μg ml−1.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 37, Issue 5, 29 April 2005, Pages 893-898