کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10559467 969627 2011 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
On-line coupling of automatic solid-phase extraction and HPLC for determination of carotenoids in serum
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
On-line coupling of automatic solid-phase extraction and HPLC for determination of carotenoids in serum
چکیده انگلیسی
The automated method developed for the determination of carotenoids uses 200 μL of serum, which was mixed with 400 μL of tetrahydrofuran, vortexed for 1 min, settled for 10 min, centrifuged for 6 min and the supernatant injected into an automatic solid-phase extraction (SPE) system for cleanup-preconcentration. A 10% water-acetonitrile mobile phase at 1.5 mL min−1 eluted the retained compounds and transferred them on-line to a reversed-phase analytical column for individual separation of the target analytes. Visible detection was performed at 450 and 460 nm. The detection limits for the target analytes were between 3 and 30 ng mL−1; the precision (expressed as relative standard deviation) ranged between 2.83 and 5.06% for repeatability and between 3.80 and 7.40% for within laboratory reproducibility. The total analysis time was 18 min. The proposed method is reliable, robust, and has an excellent potential for high-throughput use in both clinical and research laboratories.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 85, Issue 4, 30 September 2011, Pages 1842-1847
نویسندگان
, , , ,