Radiation dose around a PET scanner installation: Comparison of Monte Carlo simulations, analytical calculations and experimental results
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Radiation dose around a PET scanner installation: Comparison of Monte Carlo simulations, analytical calculations and experimental results
چکیده انگلیسی
The validation tests performed showed a satisfactory agreement between the simulated and expected results. The simulated dose rates incident on, and transmitted by the walls in our model of PET scanner room, are generally in good agreement with analytical estimates performed using the AAPM Publication No. 108 method. This provides an independent confirmation of AAPM's approach. Even in this specific field of application, GEANT4 proved to be a relevant and accurate tool for dosimetry estimates, shielding evaluation and for general radiation protection use.