Low contrast detectability performance of model observers based on CT phantom images: kVp influence
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
تشعشع
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
Object contrast was affected by kV, with differences up to 17% between the lowest and highest kV. Both models overestimated human performance and were corrected by efficiency and internal noise factors. The NPWE model reproduced better the human PC values trends showing Pearson's correlation coefficients â¥0.976 (0.954-0.987, 95% CI) for both experiments, whereas for CHO they were â¥0.706 (0.493-0.839). Bland-Altman plots showed better agreement between NPWE and humans being the average difference Î and the range of the differences α2Ï (Ï, standard deviation) of Î=â0.3%, α2Ï = [â4.0%,4.5%]. For CHO, Î=â1.2%, α 2Ï= [â10.7%,8.3%]. The NPWE model can be a useful tool to predict human performance in CT low contrast detection tasks in a standard phantom and be potentially used in protocol optimization based on kV selection.