Helical Tomotherapy of the breast: Can thermoplastic immobilization improve the reproducibility of the treatment setup and the accuracy of the delivered dose?
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Helical Tomotherapy of the breast: Can thermoplastic immobilization improve the reproducibility of the treatment setup and the accuracy of the delivered dose?
چکیده انگلیسی
Analysis of shift-roll data shows no improvement in the group of patients immobilized with the addition of thermoplastic mask. Van Herk's margin is quite large (7-10 mm) in both groups evidencing the need to perform daily setup correction. The adapted dose distribution complies well with the planned one, showing that if MVCT is used before each treatment fraction, a 3 mm margin (setup component) for CTVs expansion could be adequate.