Direct determination of sodium, potassium, chromium and vanadium in biodiesel fuel by tungsten coil atomic emission spectrometry
ترجمه فارسی عنوان
تعیین مستقیم سدیم، پتاسیم، کروم و وانادیوم در سوخت بیودیزل با طیف سنجی انتشار اتمی تنگستن
کلمات کلیدی
سوخت بیودیزل، سیم پیچ تنگستن، تجزیه و تحلیل مستقیم، انتشار اتمی، تجزیه و تحلیل فلز ردیابی، شستشو و شستشو
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• Direct analysis of biodiesel on a tungsten coil atomizer.
• Determination of Na, K, Cr and V by tungsten coil atomic emission spectrometry.
• Sample dilution with methanol or ethanol.
• Ten-microliter sample aliquots and limits of detection between 20 and 90 μg kg−1.
• Low consumption of reagents, samples and gases in a 140 s per run procedure.

High levels of sodium and potassium can be present in biodiesel fuel and contribute to corrosion, reduced performance and shorter engine lifetime. On the other hand, trace amounts of chromium and vanadium can increase the emission of pollutants during biodiesel combustion. Sample viscosity, immiscibility with aqueous solutions and high carbon content can compromise biodiesel analyzes. In this work, tungsten filaments extracted from microscope light bulbs are used to successively decompose biodiesel's organic matrix, and atomize and excite the analytes to determine sodium, potassium, chromium and vanadium by tungsten coil atomic emission spectrometry (WCAES). No sample preparation other than simple dilution in methanol or ethanol is required. Direct analysis of 10-μL sample aliquots using heating cycles with less than 150 s results in limits of detection (LOD) as low as 20, 70, 70 and 90 μg kg−1 for Na, K, Cr and V, respectively. The procedure's accuracy is checked by determining Na and K in a biodiesel reference sample and carrying out spike experiments for Cr and V. No statistically significant differences were observed between reference and determined values for all analytes at a 95% confidence level. The procedure was applied to three different biodiesel samples and concentrations between 6.08 and 95.6 mg kg−1 for Na and K, and between 0.22 and 0.43 mg kg−1 for V were obtained. The procedure is simple, fast and environmentally friendly. Small volumes of reagents, samples and gases are used and no residues are generated. Powers of detection are comparable to other traditional methods.