Comparison of the performance of three ion mobility spectrometers for measurement of biogenic amines
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Comparison of the performance of three ion mobility spectrometers for measurement of biogenic amines
چکیده انگلیسی

The performance of three different types of ion mobility spectrometer (IMS) devices: GDA2 with a radioactive ion source (Airsense, Germany), UV-IMS with a photo-ionization source (G.A.S. Germany) and VG-Test with a corona discharge source (3QBD, Israel) was studied. The gas-phase ion chemistry in the IMS devices affected the species formed and their measured reduced mobility values. The sensitivity and limit of detection for trimethylamine (TMA), putrescine and cadaverine were compared by continuous monitoring of a stream of air with a given concentration of the analyte and by measurement of headspace vapors of TMA in a sealed vial. Preprocessing of the mobility spectra and the effectiveness of multivariate curve resolution techniques (MCR-LASSO) improved the accuracy of the measurements by correcting baseline effects and adjusting for variations in drift time as well as enhancing the signal to noise ratio and deconvolution of the complex data matrix to their pure components. The limit of detection for measurement of the biogenic amines by the three IMS devices was between 0.1 and 1.2 ppm (for TMA with the VG-Test and GDA, respectively) and between 0.2 and 0.7 ppm for putrescine and cadaverine with all three devices. Considering the uncertainty in the LOD determination there is almost no statistically significant difference between the three devices although they differ in their operating temperature, ionization method, drift tube design and dopant chemistry. This finding may have general implications on the achievable performance of classic IMS devices.

The response to different amounts of TMA (in μg) that were placed in a headspace vial as a function of time for the VG-Test (top) and the GDA (bottom). Note that the ratio [TMA/(TMA + TEP)] (top) and [TMA/(TMA + RIP)] (bottom) and the clearance time increase with the amount of TMA deposited in the vial.Figure optionsDownload as PowerPoint slideHighlights
► First comparison of performance of IMS devices.
► Gas-phase ion chemistry affected by operational parameters.
► Limits of detection quite similar despite differences in devices.
► LODs determined in controlled continuous flow and in headspace vapor.
► Exponential dilution of headspace studies.