This paper presents an alternative analytical method employing energy dispersive X-ray fluorescence (EDXRF) to determine copper, iron, nickel and zinc ions in ethanol fuel samples after a pre-concentration procedure. Our pre-concentration strategy utilizes analyte retention on cation exchange chromatography paper, a convenient substrate for direct EDXRF measurements. The repeatability, expressed in terms of RSD of standard solutions containing 0.25 μg mL−1 of Cu, Fe, Ni and Zn, and calculated from fifteen consecutive measurements, was 2.5, 2.8, 3.0, and 2.7%, respectively. The limits of detection (LOD), defined as the analyte concentration that gives a response equivalent to three times the standard deviation of the blank (n = 10), were found to be 13, 15, 15 and 12 μg L−1 for Cu, Fe, Ni and Zn, respectively. The proposed method was applied to Cu, Fe, Ni and Zn determination in hydrated ethanol fuel samples collected from different gas stations.
Figure optionsDownload as PowerPoint slideHighlights
► A procedure is proposed for analyzing metal ions in ethanol fuel samples by EDXRF after a preconcentration step.
► The preconcentration process involves analyte retention on chromatography paper.
► The method does not require the sample to be subjected to time-consuming pretreatment.
Journal: Analytica Chimica Acta - Volume 722, 13 April 2012, Pages 29–33