The recent findings of antibiotic residues in aquatic environment at trace level have gained much concern for the detrimental effect on ecological and human health due to bacterial resistance. Here, the feasibility of using electrospray ionization ion mobility spectrometry (ESI-IMS) for analysis antibiotics in liquid sample is demonstrated. Reduced mobilities and collision cross sections of 18 antibiotics are experimentally measured and compared with theoretical values according to mass-mobility correlation. Gentamicin is used as an example to investigate the capability of ESI-IMS for multi-component analysis of antibiotics. Mixtures of antibiotics at different concentrations are analyzed. The estimated detection limit for amoxicillin is 0.7 mg L−1 (70 pg) and the linear range of response maintains over two orders. This method will be a potential technique for the analysis of antibiotics in aquatic environment.
Figure optionsDownload as PowerPoint slideHighlights
► The reduced mobilities of 18 antibiotics are determined.
► Establishing antibiotic mass-mobility correlation using (12,4) potential model.
► Multi-component characteristics of antibiotics can be revealed using ESI-IMS.
► Most mixtures of antibiotics can be analyzed using ESI-IMS.
► The detection limit of amoxicillin is 70 pg.
Journal: Analytica Chimica Acta - Volume 720, 30 March 2012, Pages 97–103