This paper reports a lab-on-a-chip device that counts the number of bacteria flowing through a microchannel. The bacteria number counting is realized by a microfluidic differential Resistive Pulse Sensor (RPS). By using a single microfluidic channel with two detecting arm channels placed at the two ends of the sensing section, the microfluidic differential RPS can achieve a high signal-to-noise ratio. This method is applied to detect and count bacteria in aqueous solution. The detected RPS signals amplitude for Pseudomonas aeruginosa ranges from 0.05 V to 0.17 V and the signal-to-noise ratio is 5–17. The number rate of the bacteria flowing through the sensing gate per minute is a linear function of the sample concentration. Using this experimentally obtained correlation curve, the concentration of bacteria in the sample solution can be evaluated within several minutes by measuring the number rate of the bacteria flowing through the sensing gate of this microfluidic differential RPS chip. The method described in this paper is simple and automatic, and have wide applications in determining the bacteria and cell concentrations for microbiological and other biological applications.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 681, Issues 1–2, 29 November 2010, Pages 82–86