کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1167098 960527 2011 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Optimised determinations of water in ethanol by encoded photometric near-infrared spectroscopy: A special case of sequential standard addition calibration
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Optimised determinations of water in ethanol by encoded photometric near-infrared spectroscopy: A special case of sequential standard addition calibration
چکیده انگلیسی

A special limiting case of sequential standard addition calibration (S-SAC) has been applied to measurement of the water content of ethanol using encoded photometric near infrared spectroscopy. The method has shown good comparability with certified reference materials and to measurements made by Karl Fischer titration. The technique is quick and easy to use and should have application in high throughput and process measurement, for instance in biofuels analysis at port-of-entry or in bio-refineries. The characteristics of this limiting case of S-SAC have been fully described, and the corrections required to the value obtained by extrapolation to avoid bias have been calculated. The precision of the S-SAC procedure has been studied, and proposals have been made to optimise this with respect to the analytical precision. The technique should be applicable for the measurement of water in ethanol mass fractions of up to 0.1 g g−1 with an expanded uncertainty of less than 2% (relative).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 690, Issue 1, 25 March 2011, Pages 47–52
نویسندگان
, , ,