کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1170010 | 960661 | 2008 | 7 صفحه PDF | دانلود رایگان |
A radiochemical method for the isolation of plutonium-isotopes from environmental samples, based on the use of specific extraction chromatography resins for actinides (TEVA®, Eichrom Industries, Inc.), has been set up in our laboratory and optimised for their posterior determination by alpha spectrometry (AS) or accelerator mass spectrometry (AMS). The proposed radiochemical method has replaced in our lab a well-established one based on the use of a relatively un-specific anion-exchange resin (AG® 1X8, Bio-rad Laboratories, Inc.), because it is clearly less time consuming, reduces the amounts and molarities of acid wastes produced, and reproducibly gives high radiochemical yields.In order to check the reliability of the proposed radiochemical method for the determination of plutonium-isotopes in different environmental matrixes, twin aliquots of a set of samples were prepared with TEVA® and with AG® 1X8 resins and measured by AS. Some samples prepared with TEVA® resins were measured as well by AMS. As it is shown in the text, there is a comfortable agreement between AS and AMS, which adequately validates the method.
Journal: Analytica Chimica Acta - Volume 606, Issue 2, 14 January 2008, Pages 239–245