کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1170023 960662 2007 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Determination of dissolved sulphides in waste water samples by flow-through stripping chronopotentiometry with a macroporous mercury-film electrode
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of dissolved sulphides in waste water samples by flow-through stripping chronopotentiometry with a macroporous mercury-film electrode
چکیده انگلیسی
Sulphides in water samples were determined by stripping chronopotentiometry in a computer controlled flow system with a flow-through electrochemical cell. The working electrode was a porous glassy carbon electrode coated with Nafion and mercury. The sample was diluted with 0.1 mol L−1 NaOH and analysed. Sulphides in the sample were collected in the porous electrode as mercury sulphide and then stripped by a current of −500 μA. The limit of detection was found to be 1.6 μg L−1 and 0.5 μg L−1 for 1 mL and 5 mL of preconcentrated sample, respectively. The linear range for 1 mL sample was found to be 5-400 μg L−1. The repeatability and reproducibility was found to be 2.6% and 4.8%, respectively. The method was applied to analyses of waste water samples from a tannery.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 588, Issue 1, 4 April 2007, Pages 16-19
نویسندگان
, , , , ,