کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1170565 960684 2007 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Extracting syringe for extraction of phthalate esters in aqueous environmental samples
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Extracting syringe for extraction of phthalate esters in aqueous environmental samples
چکیده انگلیسی
Toluene was the optimal solvent to use for extraction. Due to the large variation in polarity of phthalate esters, 50% methanol as organic modifier had to be added to the samples so as to extract the most nonpolar phthalate esters; di-2-ethylhexylphthalate and di-n-octylphthalate, whereas the other four relatively polar phthalate esters were extracted from unmodified samples. No significant difference between extraction of river water, leachate water from a landfill and reagent water was noted, except for minor deviations. The extraction time was 20 min for extraction of a 1-mL sample, resulting in a good linearity for all aqueous media investigated, good enrichment factors (54-110 folds) and low LOD values (0.2-10 ng mL−1) and relative standard deviation (%R.S.D.; 0.9-3.7%).
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 594, Issue 2, 2 July 2007, Pages 240-247
نویسندگان
, , , , ,