کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1170591 960685 2007 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Colour quenching corrections on the measurement of 90Sr through Cerenkov counting
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Colour quenching corrections on the measurement of 90Sr through Cerenkov counting
چکیده انگلیسی

The determination of 90Sr through the Cerenkov radiation emitted by its descendant 90Y is a well-known method and firmly established in literature. Nevertheless, in order to obtain an accurate result based on a Cerenkov measurement, the experimental work must be extremely rigorous because the efficiency of Cerenkov counting is especially sensitive to the presence of colour. Any traces of colour in the sample produce a decrease in the number of photons detected in the photomultipliers and, therefore, this might cause a diminution in Cerenkov counting efficiency. It is essential not only to detect the effect of colour quenching in the sample but also to correct the decrease in counting efficiency. For this reason, colour quenching correction curves versus counting efficiency are usually done when measuring through Cerenkov counting. One of the most widely used techniques to evaluate colour quenching in these measurements is the channel ratio method, which consists of the measurement of the shift of the spectrum measuring the ratio of counts in two different windows. The selection of the windows for the application of the corrections might have an influence on the quality of the fitting parameters of the correction curves efficiency versus colour quenching degree and hence on the final 90Sr result. This work is focused on the calculation of the counting efficiency decrease using the channel ratio method and on obtaining the best fitting correction curve. For this purpose, empirical curves obtained through artificial quenchers have been studied and the results have been tested in real samples. Additionally, given that the Packard Tri-Carb 3170 TR/SL liquid scintillation counter is a novel detector for use in Cerenkov counting, the previous calibration of the Tri-Carb 3170 TR/SL detector, necessary for the measurement of 90Sr, is included.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 604, Issue 2, 5 December 2007, Pages 184–190
نویسندگان
, , , ,