A pulsed valve was used to increase the pressure within the trapping region of a low-pressure linear ion trap by situating the pulsed valve close to the ion trapping region. The pressure was estimated to increase from a background pressure of 3.5e–5 Torr of nitrogen to 0.49 mTorr at the center of the trap. The increased pressure allowed excitation periods to be reduced from 100 to 25 ms without suffering losses in MS/MS efficiency during dipolar excitation. The reduction in excitation period translates into an increase in the overall duty cycle of the scan.
Graphical AbstractIn-trap fragmentation efficiencies and scan duty cycle are improved using a pulsed valve to increase local pressure within low pressure linear ion trap.Figure optionsDownload high-quality image (79 K)Download as PowerPoint slide
Journal: Journal of the American Society for Mass Spectrometry - Volume 20, Issue 9, September 2009, Pages 1714–1717