Performance Evaluation of a High-field Orbitrap Mass Analyzer
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Performance Evaluation of a High-field Orbitrap Mass Analyzer
چکیده انگلیسی

A new design of the Orbitrap™ mass analyzer is presented. Higher frequencies of ion oscillations and hence higher resolving power over fixed acquisition time are achieved by decreasing the gap between the inner and outer Orbitrap electrodes, thus providing higher field strength for a given voltage. Experimental results confirm maximum FWHM resolving power in excess of 350,000 at m/z 524 and 600,000 at m/z 195, isotopic resolution of proteins above 40 kDa, and a single-shot dynamic range of 25,000. It was also found that mass shifts in the new design depend very little on space charge inside the analyzer. This performance was achieved using higher voltages and by careful balancing of construction tolerances and operation parameters, which appeared to vary in narrower ranges of tuning than for a standard Orbitrap analyzer.

Graphical AbstractThis work is devoted to experimental characterization of a novel, high-field Orbitrap mass analyzer with a decreased gap between electrodes.Figure optionsDownload high-quality image (191 K)Download as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of the American Society for Mass Spectrometry - Volume 20, Issue 8, August 2009, Pages 1391–1396