کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1195819 | 964415 | 2006 | 6 صفحه PDF | دانلود رایگان |

This paper describes a novel emitter tip having the shape of a nib and based on an open structure for nano-electrospray ionization mass spectrometry (nanoESI-MS). The nib structure is fabricated with standard lithography techniques using SU-8, an epoxy-based negative photoresist. The tip is comprised of a reservoir, a capillary slot and a point-like feature, and is fabricated on a silicon wafer. We present here a novel scheme for interfacing such nib tips to MS by applying the ionization voltage directly onto the semi-conductor support. The silicon support is in direct contact with the liquid to be analyzed at the reservoir and microchannel level, thus allowing easy use in ESI-MS. This scheme is especially advantageous for automated analysis as the manual step of positioning a metallic wire into the reservoir is avoided. In addition, the analysis performance was enhanced compared with the former scheme, as demonstrated by the tests of standard peptides (gramicidin S, Glu-fibrinopeptide B). The limit of detection was determined to be lower than 10−2 μM. Due to their enhanced performance, these microfabricated sources might be of great interest for analysis requiring very high sensitivity, such as proteomics analysis using nanoESI-MS.
Journal: Journal of the American Society for Mass Spectrometry - Volume 17, Issue 1, January 2006, Pages 75–80