Evaluation of the impact of matrix effect on quantification of pesticides in foods by gas chromatography–mass spectrometry using isotope-labeled internal standards
ترجمه فارسی عنوان
بررسی تأثیر تاثیر ماتریکس بر اندازه گیری آفت کش ها در غذاها با استفاده از اسپکترومتری جرمی کروماتوگرافی گاز با استفاده از استانداردهای داخلی نشانه ایزوتوپ
کلمات کلیدی
اثر ماتریکس، طیف سنجی جرمی ایزوتوپ-رقت. استاندارد داخلی با برچسب دتریوم، آفت کش ها، مطابقت ماتریکس، منحنی کالیبراسیون
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• Matrix effect in GC–MS evaluated using isotope-labeled internal standards.
• Matrix-free calibration solution results were biased from spiked concentrations.
• Calibration curve slope was influenced by the matrix in calibration solution.
• Matrix-matching was required for very accurate quantification, even with isotope-labeled internal standards.

The impact of the matrix effect in GC–MS quantification of pesticides in food using the corresponding isotope-labeled internal standards was evaluated. A spike-and-recovery study of nine target pesticides was first conducted using paste samples of corn, green soybean, carrot, and pumpkin. The observed analytical values using isotope-labeled internal standards were more accurate for most target pesticides than that obtained using the external calibration method, but were still biased from the spiked concentrations when a matrix-free calibration solution was used for calibration. The respective calibration curves for each target pesticide were also prepared using matrix-free calibration solutions and matrix-matched calibration solutions with blank soybean extract. The intensity ratio of the peaks of most target pesticides to that of the corresponding isotope-labeled internal standards was influenced by the presence of the matrix in the calibration solution; therefore, the observed slope varied. The ratio was also influenced by the type of injection method (splitless or on-column). These results indicated that matrix-matching of the calibration solution is required for very accurate quantification, even if isotope-labeled internal standards were used for calibration.