Analytical scale active flow technology first generation silica monolithic columns kitted out in curtain flow mode of operation were studied for the first time. A series of tests were undertaken assessing the column efficiency, peak asymmetry and detection sensitivity. Two curtain flow columns were tested, one with a fixed outlet ratio of 10% through the central exit port, the other with 30%. Tests were carried out using a wide range in inlet flow segmentation ratios. The performance of the curtain flow columns were compared to a conventional monolithic column. The gain in theoretical plates achieved in the curtain flow mode of operation was as much as 130%, with almost Gaussian bands being obtained. Detection sensitivity increased by as much as 250% under optimal detection conditions. The permeability advantage of the monolithic structure together with the active flow technology makes it a priceless tool for high throughput, sensitive, low detection volume analyses.
Journal: Journal of Chromatography A - Volume 1351, 18 July 2014, Pages 56–60