Harmonization of the quantitative determination of volatile fatty acids profile in aqueous matrix samples by direct injection using gas chromatography and high-performance liquid chromatography techniques: Multi-laboratory validation study
ترجمه فارسی عنوان
هماهنگی کمی تعیین پروتئین های اسید چرب فرار در نمونه های ماتریکس آب با تزریق مستقیم با استفاده از کروماتوگرافی گاز و تکنیک های کروماتوگرافی مایع با عملکرد بالا: مطالعه اعتبار چند آزمایشگاهی
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• A multi-laboratory validation for VFAs measurement is described.
• Acceptance criteria for performance parameters were suggested.
• Instrument precision should be lower than 1.5% RSD.
• Calibration linearity should be assessed using RSDSENSITIVITY and %RE.
• Appropriate VFAs measurements: precision (RSD ≤ 1.5%) and recovery (97–103%).

The performance parameters of volatile fatty acids (VFAs) measurements were assessed for the first time by a multi-laboratory validation study among 13 laboratories. Two chromatographic techniques (GC and HPLC) and two quantification methods such as external and internal standard (ESTD/ISTD) were combined in three different methodologies GC/ESTD, HPLC/ESTD and GC/ISTD. Linearity evaluation of the calibration functions in a wide concentration range (10–1000 mg/L) was carried out using different statistical parameters for the goodness of fit. Both chromatographic techniques were considered similarly accurate. The use of GC/ISTD, despite showing similar analytical performance to the other methodologies, can be considered useful for the harmonization of VFAs analytical methodology taking into account the normalization of slope values used for the calculation of VFAs concentrations. Acceptance criteria for VFAs performance parameters of the multi-laboratory validation study should be established as follows: (1) instrument precision (RSDINST ≤ 1.5%); (2) linearity (R2 ≥ 0.998; RSDSENSITIVITY ≤ 4%; REMAX ≤ 8%; REAVER ≤  3%); (3) precision (RSD ≤ 1.5%); (4) trueness (recovery of 97–103%); (5) LOD (≤3 mg/L); and (6) LOQ (10 mg/L).