کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1203192 | 965102 | 2011 | 11 صفحه PDF | دانلود رایگان |

The peak parking (PP) method probes the longitudinal diffusion coefficient of a compound at a single location along the chromatographic column. We extended to a so-called multi-location peak parking (MLPP) method, in which a large number of axial locations along the column are selected in order to check the validity of the conventional PP method and to reveal possible defaults in the structure of the packed bed or pitfalls of the PP and the MLPP methods. MLPP was applied to a series of HILIC columns, including a 5.0 μm Venusil, a 3.0 μm Luna-diol, three 2.7 μm Halo, and a 1.7 μm Kinetex columns. The results demonstrate that the MLPP method may reveal local heterogeneities in the axial diffusion of small retained low molecular weight compounds along the column. Most importantly, experiments show that the sample zone should not be parked in the entrance of the column (i.e., at <1/10 th of the column length). The abrupt drop in the flow rate considerably affects the peak shape and prevents scientists from using the conventional PP method. Practical solutions to cope with that problem are proposed and their success/failure are discussed.
Journal: Journal of Chromatography A - Volume 1218, Issue 7, 18 February 2011, Pages 896–906