کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1205404 965195 2008 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Impact of an error in the column hold-up time for correct adsorption isotherm determination in chromatography: II. Can a wrong column porosity lead to a correct prediction of overloaded elution profiles?
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Impact of an error in the column hold-up time for correct adsorption isotherm determination in chromatography: II. Can a wrong column porosity lead to a correct prediction of overloaded elution profiles?
چکیده انگلیسی

The adsorption isotherm was determined for phenol in methanol/water on a C-8 stationary phase using frontal analysis in staircase mode, assuming different total column porosities, from 1 to 87%. Each set of adsorption isotherm data, with a certain column porosity, was fitted to various adsorption models and the generated parameters were used to calculate overloaded elution band profiles that were compared with experiments. It was found that the bi-Langmuir model had an optimum fit for a porosity that corresponds well with the value found experimentally. The adsorption energy distribution (AED) calculations and error analysis confirmed a bimodal energy distribution. It was also found that band profiles can be accurately predicted with a quite arbitrary chosen porosity, under prerequisite that a wrong but flexible adsorption model is chosen instead of the correct one. The latter result is very useful for quick optimizations of preparative separations where the exact value of the column porosity is not available.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1194, Issue 2, 20 June 2008, Pages 205–212
نویسندگان
, , , , ,