A study of the parameters affecting the accuracy of the total pore blocking method
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A study of the parameters affecting the accuracy of the total pore blocking method
چکیده انگلیسی

We report on a study wherein we investigate the different factors affecting the accuracy of the total pore blocking method to determine the interstitial volume of reversed-phase packed bed columns. Octane, nonane, decane and dodecane were all found to be suitable blocking agents, whereas heptane already dissolves too well in the applied fully aqueous buffers. The method of moments needs to be used to accurately determine the elution times, and a proper correction for the frit volume is needed. Failing to do so can lead to errors on the observed interstitial volume of the order of 2% or more. It has also been shown that the application of a high flow rate or a high pressure does not force the blocking agent out of the mesopores of the particles. The only potential source of loss of blocking agent is dissolution into the mobile phase (even though this is a buffered fully aqueous solution). This effect however only becomes significant after the elution of 400 geometrical column volumes, i.e., orders more than needed for a regular total pore blocking experiment.