کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1209022 1493753 2007 16 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
The bandwidth in gradient elution chromatography with a retained organic modifier
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
The bandwidth in gradient elution chromatography with a retained organic modifier
چکیده انگلیسی

The variance of a chromatographic band is derived in the case of RPLC gradient elution when the organic modifier is significantly retained onto the stationary phase. This derivation is based on the extension of a model due to Poppe et al. [H. Poppe, J. Paanakker, M. Bronckhorst, J. Chromatogr., 204 (1981) 77] which assumes that the gradient front remains unchanged and propagates along the column at the same speed as the mobile phase, following piston flow. Theoretical and experimental results are compared in the case of caffeine on a C1-silica stationary phase eluted with an acetonitrile gradient. The actual retention behaviors of caffeine and acetonitrile were implemented in the theoretical calculations. The model predicts compression factors between 0.71 and 0.34 for relatively smooth gradient steepness, βt0βt0, between 0.009 and 0.054 while the corresponding experimental band compression factors vary between 1.01 and 0.43 for the very same gradient steepness. The model underestimation of these factors arises likely from the strong deviation of the actual retention behavior from the prediction of the Linear Solvent Strength Model (LSSM).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1145, Issues 1–2, 23 March 2007, Pages 67–82
نویسندگان
, ,