کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1209032 | 1493753 | 2007 | 6 صفحه PDF | دانلود رایگان |

The suitability of the Eksigent Express 800 microfluidic eight-channel HPLC instrument for multiparallel normal-phase chiral analysis in support of high-throughput pharmaceutical process research was investigated. Analysis of test mixtures containing the two enantiomers of benzoin and the closely related (R,S)-dihydrobenzoin, was carried out in a 96-well microplate, affording rapid (<2 h) and accurate assessment of enantiopurity. In a second example, use of the instrument to support high-throughput catalyst screening of the asymmetric hydrogenation of a prochiral unsaturated ester is presented, in which method development (gradient screening of four columns and two eluents, followed by optimization to afford a fast analytical method) and analysis of a 96-well microplate was carried out within a single working day. This represents a considerable improvement over conventional analysis techniques that usually take several days to complete.
Journal: Journal of Chromatography A - Volume 1145, Issues 1–2, 23 March 2007, Pages 149–154