عنوان انگلیسی مقاله ISI
A fast and simple assay for busulfan in serum or plasma by liquid chromatography-tandem mass spectrometry using turbulent flow online extraction technology
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
busulfanLC–MS - LC-MSLC–MS/MS - LC-MS / MSTurbulent flow - جریان آشفتهliquid chromatography–mass spectrometry - طیف سنجی جرم کروماتوگرافی مایعTandem mass spectrometry - طیف سنجی جرمی پشت سر هم یا متوالیliquid chromatography - کروماتوگرافی مایعhigh performance liquid chromatography - کروماتوگرافی مایع با کارایی بالاliquid chromatography–tandem mass spectrometry - کروماتوگرافی مایع و اسپکترومتری توده دو طرفهHPLC - کروماتوگرافی مایعی کارا
موضوعات مرتبط
مهندسی و علوم پایه
شیمی
شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
![عکس صفحه اول مقاله: A fast and simple assay for busulfan in serum or plasma by liquid chromatography-tandem mass spectrometry using turbulent flow online extraction technology A fast and simple assay for busulfan in serum or plasma by liquid chromatography-tandem mass spectrometry using turbulent flow online extraction technology](/preview/png/1213660.png)
چکیده انگلیسی
Busulfan is used in myeloablative preparation regimens for hematopoietic bone marrow transplantation. Due to its narrow therapeutic range therapeutic drug monitoring of busulfan is recommended. In this study a fast and simple method for measuring busulfan in serum or plasma by liquid chromatography-tandem mass spectrometry (LC-MS/MS) has been developed utilizing turbulent flow online extraction technology. Serum or plasma was mixed with acetonitrile containing d8-busulfan. After centrifugation the supernatant was injected onto a turbulent flow preparatory column then transferred to a C18 analytical column monitored by a tandem mass spectrometer set at positive electrospray ionization. The analytical cycle time was 4.0 min. The method was linear from 0.15 to 41.90 μmol/L with an accuracy of 87.9-103.0%. Inter- and intra-assay CVs across four concentration levels were 2.1-7.8%. No significant carryover or ion suppression was observed. No interference was observed from commercial control materials containing more than 100 compounds. Comparison with a well established LC-MS/MS method using patient specimens (n = 45) showed a mean bias 1.3% with Deming regression of slope 1.02, intercept â0.02 μmol/L, and a linear correlation coefficient 0.9883. The LC-MS/MS method coupled with turbulent flow online sample cleaning technology described here offers reliable busulfan quantitation in serum or plasma with minimum manual sample preparation and was fully validated for clinical use.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 878, Issue 31, 1 December 2010, Pages 3255-3258
Journal: Journal of Chromatography B - Volume 878, Issue 31, 1 December 2010, Pages 3255-3258
نویسندگان
Dustin R. Bunch, Courtney Heideloff, James C. Ritchie, Sihe Wang,