A liquid chromatography/tandem mass spectrometry method for the simultaneous quantification of valsartan and hydrochlorothiazide in human plasma
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A liquid chromatography/tandem mass spectrometry method for the simultaneous quantification of valsartan and hydrochlorothiazide in human plasma
چکیده انگلیسی

A rapid and sensitive liquid chromatography/tandem mass spectrometry (LC/MS/MS) method was developed and validated for simultaneous quantification of valsartan and hydrochlorothiazide in human plasma. After a simple protein precipitation using acetonitrile, the analytes were separated on a Zorbax SB-Aq C18 column using acetonitrile −10 mM ammonium acetate (60:40, v/v, pH 4.5) as mobile phase at a flow rate of 1.2 mL/min. Valsartan and hydrochlorothiazide were eluted at 2.08 min and 1.50 min, respectively, ionized using ESI source, and then detected by multiple reaction monitoring (MRM) mode. The precursor to product ion transitions of m/z 434.2–350.2 and m/z 295.9–268.9 were used to quantify valsartan and hydrochlorothiazide, respectively. The method was linear in the concentration range of 4–3600 ng/mL for valsartan and 1–900 ng/mL for hydrochlorothiazide. The method was successfully employed in a pharmacokinetic study after an oral administration of a dispersible tablet containing 80 mg valsartan and 12.5 mg hydrochlorothiazide to each of the 20 healthy volunteers.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 852, Issues 1–2, 1 June 2007, Pages 436–442
نویسندگان
, , , , , , ,