کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1216671 1494192 2006 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
High-performance liquid chromatographic assay with UV detection for measurement of dihydrouracil/uracil ratio in plasma
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
High-performance liquid chromatographic assay with UV detection for measurement of dihydrouracil/uracil ratio in plasma
چکیده انگلیسی

A rapid, robust and sensitive HPLC method for analysis of uracil (U) and dihydrouracil (UH2) in plasma was developed using solid phase extraction and ultraviolet detection. Separation was achieved with a SymmetryShield RP18 column and an Atlantis dC18 column using a 10 mM potassium phosphate buffer as mobile phase. Compounds were eluted within 15 min without interference. Recovery was 80.4 and 80.6% for U and UH2. Calibration curves were linear from 2.5 to 80 ng/mL for U and 6.75 to 200 ng/mL for UH2. The LLQ was, respectively, 2.5 ng/mL for U, and 6.75 ng/mL for UH2. Within-run and between-run precision were less than 5.94% and inaccuracy did not exceed 7.80%. The overall procedure has been applied to correlate UH2/U ratio with dihydropyrimidine dehydrogenase activity in 165 cancer patients.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 834, Issues 1–2, 13 April 2006, Pages 170–177