![عکس صفحه اول مقاله: Increasing conclusiveness of clinical breath analysis by improved baseline correction of multi capillary column – ion mobility spectrometry (MCC-IMS) data Increasing conclusiveness of clinical breath analysis by improved baseline correction of multi capillary column – ion mobility spectrometry (MCC-IMS) data](/preview/png/1220925.png)
• Three baseline correction methods are evaluated in the analysis of MCC-IMS data.
• Top-hat method is selected as the optimum baseline correction method.
• Top-hat method is included in the data analysis strategy for MCC-IMS datasets.
• Large dataset on a healthy and respiratory disease population is analyzed.
• MCC-IMS spectra differences associated with respiratory diseases are provided.
Current challenges of clinical breath analysis include large data size and non-clinically relevant variations observed in exhaled breath measurements, which should be urgently addressed with competent scientific data tools. In this study, three different baseline correction methods are evaluated within a previously developed data size reduction strategy for multi capillary column – ion mobility spectrometry (MCC-IMS) datasets. Introduced for the first time in breath data analysis, the Top-hat method is presented as the optimum baseline correction method. A refined data size reduction strategy is employed in the analysis of a large breathomic dataset on a healthy and respiratory disease population. New insights into MCC-IMS spectra differences associated with respiratory diseases are provided, demonstrating the additional value of the refined data analysis strategy in clinical breath analysis.
Figure optionsDownload as PowerPoint slide
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 127, 5 August 2016, Pages 170–175