An HPLC/UV method was first optimized for the separation and quantitation of human acylated and unacylated (or des-acyl) ghrelin from aqueous solutions. This method was validated by an original approach using accuracy profiles based on tolerance intervals for the total error measurement. The concentration range that achieved adequate accuracy extended from 1.85 to 59.30 μM and 1.93 to 61.60 μM for acylated and unacylated ghrelin, respectively. Then, optimal temperature, pH and buffer for sample storage were determined. Unacylated ghrelin was found to be stable in all conditions tested. At 37 °C acylated ghrelin was stable at pH 4 but unstable at pH 7.4, the main degradation product was unacylated ghrelin. Finally, this validated HPLC/UV method was used to evaluate the binding of acylated and unacylated ghrelin to liposomes.