چکیده انگلیسی
The main results were that centrifugation of samples reduced variability. When a batch of 24 samples was analysed, the use of crimp caps reduced evaporation. Samples in glass vials with crimp caps could be stored in a refrigerator for up to 14 days without large variability in concentration compared to plastic vials which demonstrated variability already when stored for more than 1 day. We conclude that variability in microdialysis results can occur in relation to storage and analysis routines if routines are not optimised concerning evaporation. Centrifugation before analyses, glass vials with crimp caps even during frozen storage, and attention to minimal times for samples to be uncapped during analysis all contribute to minimise variability in the handling and analysis of microdialysis samples.