A rapid and simple method for procaine determination was developed by flow injection analysis (FIA) using a screen-printed carbon electrode (SPCE) as amperometric detector. The present method is based on the amine/hydroxylamine oxidation from procaine monitored at 0.80 V on SPCE in sodium acetate solution pH 6.0. Using the best experimental conditions assigned as: pH 6.0, flow rate of 3.8 mL min−1, sample volume of 100 μL and analytical path of 30 cm it is possible to construct a linear calibration curve from 9.0 × 10−6 to 1.0 × 10−4 mol L−1. The relative standard deviation for 5.0 × 10−5 mol L−1 procaine (15 repetitions using the same electrode) is 3.2% and detection limit calculated is 6.0 × 10−6 mol L−1. Recoveries obtained for procaine gave a mean values from 94.8 to 102.3% and an analytical frequency of 36 injections per hour was achieved. The method was successfully applied for the determination of procaine in pharmaceutical formulation without any pre-treatment, which are in good accordance with the declared values of manufacturer and an official method based on spectrophotometric analysis.
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 43, Issue 1, 4 January 2007, Pages 315–319