A method for Cr determination in medicinal plants using direct solid sampling graphite furnace high-resolution continuum source atomic absorption spectrometry was developed. Modifiers were dispensable. Pyrolysis and atomization temperatures were 1500 °C and 2400 °C, respectively. Slopes of calibration curves (50–750 pg Cr, R2 > 0.999) using aqueous and solid standards coincides in 96%, indicated feasibility of aqueous calibration for solid sampling of medicinal plants. Accuracy was checked by analysis of four plant certified reference materials. Results were in agreement at 95% confidence level with certified and non-certified values. Ten samples of medicinal plants were analyzed and Cr contents were in the 1.3–17.7 μg g− 1 Cr range. The highest RSD (n = 5) was 15.4% for the sample Melissa officinalis containing 13.9 ± 2.1 μg g− 1 Cr. The limit of detection was 3.3 ng g− 1 Cr.
► Direct solid sampling is first time employed for Cr in plant materials.
► Calibration curves with liquids and solids are coincident.
► Microanalysis of plants for Cr is validated by reference materials.
► The proposed HR-CS GF AAS method is environmental friendly.
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volume 78, 1 December 2012, Pages 58–61