کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1239884 1495726 2012 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Ultratrace determination of lead by hydride generation in-atomizer trapping atomic absorption spectrometry: Optimization of plumbane generation and analyte preconcentration in a quartz trap-and-atomizer device
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Ultratrace determination of lead by hydride generation in-atomizer trapping atomic absorption spectrometry: Optimization of plumbane generation and analyte preconcentration in a quartz trap-and-atomizer device
چکیده انگلیسی

A compact trap-and-atomizer device and a preconcentration procedure based on hydride trapping in excess of oxygen over hydrogen in the collection step, both constructed and developed previously in our laboratory, were employed to optimize plumbane trapping in this device and to develop a routine method for ultratrace lead determination subsequently. The inherent advantage of this preconcentration approach is that 100% preconcentration efficiency for lead is reached in this device which has never been reported before using quartz or metal traps. Plumbane is completely retained in the trap-and-atomizer device at 290 °C in oxygen-rich atmosphere and trapped species are subsequently volatilized at 830 °C in hydrogen-rich atmosphere. Effect of relevant experimental parameters on plumbane trapping and lead volatilization are discussed, and possible trapping mechanisms are hypothesized. Plumbane trapping in the trap-and-atomizer device can be routinely used for lead determination at ultratrace levels reaching a detection limit of 0.21 ng ml− 1 Pb (30 s preconcentration, sample volume 2 ml). Further improvement of the detection limit is feasible by reducing the blank signal and increasing the trapping time.


► In-atomizer trapping HG-AAS was optimized for Pb.
► A compact quartz trap-and-atomizer device was employed.
► Generation, preconcentration and atomization steps were investigated in detail.
► 100% preconcentration efficiency for lead was reached.
► Routine analytical method was developed for Pb determination (LOD of 0.2 ng ml− 1 Pb).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volumes 71–72, May–June 2012, Pages 40–47
نویسندگان
,