Correlation between laser-induced breakdown spectroscopy signal and moisture content
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Correlation between laser-induced breakdown spectroscopy signal and moisture content
چکیده انگلیسی

The possibility of using Laser-Induced Breakdown Spectroscopy (LIBS) for measuring the moisture content of fresh food samples is studied. The normalized line emission of oxygen is highly correlated with the moisture content of the sample, cheese in our case, and can be used as a moisture marker in situations where oxygen interference from the matrix is not a critical issue. The linear correlation between the oxygen signal and the moisture content in the sample shows great potential for using LIBS as an alternative spectroscopic method for moisture monitoring.


► Quantitative moisture measurement by LIBS.
► Use of matrix effects and normalization for physical information on the sample.
► Use of signal from oxygen and CN radical in air background for moisture measurement.