کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1241037 | 969160 | 2006 | 18 صفحه PDF | دانلود رایگان |

Plasma-related matrix effects induced by a comprehensive list of matrix elements (a total of fifty-one matrices) in inductively coupled plasma-atomic emission spectrometry were investigated and used to confirm that matrix effects caused by elements with a low second ionization potential are more severe than those from matrix elements having a low first ionization potential. Although the matrix effect is correlated unambiguously with the second ionization potential of a matrix, the correlation is not monotonic, which suggests that at least one other factor is operative. Through study of a large pool of matrix elements, it becomes possible to identify another critical parameter that defines the magnitude of the matrix effect; namely the presence of low-lying energy levels in the doubly charged matrix ion. Penning ionization by Ar excited states is proposed as the dominant mechanism for both analyte ionization/excitation and matrix effects; matrices with a low second ionization potential can effectively quench the population of Ar excited states through successive Penning ionization followed by ion-electron recombination and lead to more severe matrix effects.
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volume 61, Issue 6, June 2006, Pages 642–659