کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1241490 969208 2006 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
The effect of pre-evaporation on ion distributions in inductively coupled plasma mass spectrometry
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
The effect of pre-evaporation on ion distributions in inductively coupled plasma mass spectrometry
چکیده انگلیسی

The connecting tube (2 or 5-mm i. d., 11-cm long) between the spray chamber and the torch was heated (to 400 °C) to investigate the effect of pre-evaporation on the distribution of ions in inductively coupled plasma mass spectrometry (ICP-MS). Axial and radial profiles of analyte ions (Al+, V+, Cr+, Ni+, Zn+, Mn+, Zn+, As+, Se+, Mo+, Cd+, Sb+, La+, Pb+) in 1% HNO3 as well as some polyatomic ions (LaO+, ArO+, ArN+, CO2+) were simultaneously obtained on a time-of-flight ICP-MS instrument. Upon heating the connecting tube, the optimal axial position of all elements shifted closer to the load coil. Without the heated tube, 3.5 mm was the compromise axial position for multielemental analysis, which was optimal for 6 analytes. With the heated tube, this position became 1.5 mm, which was then optimal for 9 of the 14 analytes. Furthermore, the radial profiles, which were wide with a plateau in their middle without heating, became significantly narrower and Gaussian-like with a heated tube. This narrowing, which was most important for the 5-mm tube, slightly (by a factor of two at the most) yet significantly (at the 95% confidence level) improved the sensitivity of all elements but Mn upon optimisation of the axial position for compromise multi-element analysis. Furthermore, a concurrent decrease in the standard deviation of the blank was significant at the 95% confidence level for 9 of the 14 analytes. For most of the analytes, this translated into a two-fold to up to an order of magnitude improvement in detection limit, which is commensurate with a reduction of noise resulting from the smaller droplets entering the plasma after traversing the pre-evaporation tube.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volume 61, Issue 2, February 2006, Pages 157–163
نویسندگان
, ,