کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1242011 1495803 2015 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Electrodialytic matrix isolation for metal cations
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Electrodialytic matrix isolation for metal cations
چکیده انگلیسی


• Cationic heavy metal ions were quantitatively transferred from sample matrices.
• Acidic acceptor solution was effective.
• No significant sample pH effects were observed.
• The approach provides better reproducibility than conventional wet digestion pretreatment.

Electrodialytic ion transfer was studied as a matrix isolation tool for heavy metal determinations. An ion transfer device (ITD) was used for the transfer of heavy metal cations. Under optimized flow rates applied voltage and receptor composition, heavy metal ions were quantitatively transferred at concentrations spanning µg L−1 to mg L−1. As long as the sample pH was acidic, there was no significant sample pH effect on the transfer efficiencies. Significant salt concentrations (>1 mM NaCl), however, decreased the transfer efficiency. This could be ameliorated (up to 5 mM NaCl) by transient instead of continuous sample introduction. The device was applied to the determination of Fe, Cu and Zn in equine and bovine serum; the reproducibility was better than conventional digestion method.

Figure optionsDownload as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 132, 15 January 2015, Pages 228–233
نویسندگان
, , , , , ,