کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1243854 969673 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Online determination of copper in aluminum alloy by microchip solvent extraction using isotope dilution ICP-MS method
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Online determination of copper in aluminum alloy by microchip solvent extraction using isotope dilution ICP-MS method
چکیده انگلیسی

Isotope dilution mass spectroscopy (IDMS)/ICP-MS combined with microchip solvent extraction was successfully applied for the online determination of copper in an aluminum alloy. The microchip solvent extraction was developed for the separation of Cu from major element, and optimal pH range was wider than that of the batchwise extraction method. The dimensions of the microchip were 30 mm × 70 mm and that of micro-channel on the microchip was 180 μm wide and 40 μm deep. The copper complex with 8-hydroxyquinoline was extracted into o-xylene at pH 5.5 and back extracted with 0.1 mol l−1 nitric acid at flow rate of 20 μl min−1. The total extraction efficiency (water/organic solvent/nitric acid) was around 40%. IDMS/ICP-MS was coupled with solvent extraction for precise determination of Cu. The extraction and back-extraction on the microchip took about 1 s and the total measurement time for the IDMS/ICP-MS was about 40 s/sample. The blank value of this method was 0.1 ng g−1. The proposed method was used for the determination of Cu in Al standard materials (JSAC 0121-C, The Japan Society for Analytical Chemistry and 7074 Al alloy, Nippon Light Metal Co. Ltd.). The obtained analytical results are in good agreement with the certified values.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 79, Issue 4, 15 September 2009, Pages 1001–1005
نویسندگان
, , , ,