کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1245325 | 969718 | 2008 | 10 صفحه PDF | دانلود رایگان |

A hydrophilic interaction high performance liquid chromatography–tandem mass spectrometric method has been developed and validated for simultaneous quantification of dacarbazine (DTIC) and its terminal metabolite, 5-amino-4-imidazole-carboxamide (AIC) in human plasma. The plasma samples are first extracted by a C8 + SCX mixed-mode 96-well plate to extend the extraction stability of DTIC and AIC. The extracted residues are further cleaned by a primary and secondary amine (PSA) adsorbent for minimization of matrix effect. Analyses are done on an Amide-80 HPLC column coupled to a tandem mass spectrometer fitted with an atmospheric pressure turbo ion spray ionization interface in the positive-ion mode. Both DTIC and AIC have reproducible retention times on the Amide-80 HPLC column. This type of column not only has an excellent column life (over 4000 injections), but also has zero carryover effect. The injection volume should be limited at 10 μL or less to avoid the peak splitting. The validated concentration ranges are from 0.5 to 500 ng/mL for DTIC and from 2.0 to 500 ng/mL for AIC. The validated method has been successfully applied to determine the pharmacokinetic profiles for human patients receiving DTIC infusions.
Journal: Talanta - Volume 77, Issue 1, 19 October 2008, Pages 412–421