کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1246396 969754 2006 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Molecularly imprinted TiO2 thin film using stable ground-state complex as template as applied to selective electrochemical determination of mercury
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Molecularly imprinted TiO2 thin film using stable ground-state complex as template as applied to selective electrochemical determination of mercury
چکیده انگلیسی

Molecular recognition sites for mercury ions were imprinted in TiO2 film using stable ground-state complex of 1-amino-8-naphtol-3,6-disodium sulfonate (ANDS) and mercury ions as template. The complex ratio between mercury ions and ANDS was estimated to be 2:1. Compared with the controlled and pure TiO2 electrodes, the imprinted electrode revealed selectivity towards the imprinted ions. Linear calibration plots for mercury ions were obtained and the regression equation was Ip (μA) = 4.29 × 10−7 + 19.40 [Hg2+] with a detection limit of 3.06 × 10−9 mol/l. The imprinted electrode could be used for more than 1 month. Recoveries were calculated at both high and low concentrations, with a mean recovery of 99.6%.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 68, Issue 4, 15 February 2006, Pages 1120–1125
نویسندگان
, , , , ,