کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1246761 969768 2009 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Cyclodextrin-MEEKC for the analysis of oxybutynin and its impurities
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Cyclodextrin-MEEKC for the analysis of oxybutynin and its impurities
چکیده انگلیسی

The development of a cyclodextrin-MEEKC method for the analysis of oxybutynin and five related impurities is described. Experimental design strategies were applied in order to reach baseline separation of the compounds in a short analysis time. Mixture design made it possible to find the best composition for the microemulsion acting as pseudostationary phase, which was constituted by 89.1% 10 mM borate buffer pH 9.2, 1.7% n-heptane, 9.2% SDS/n-butanol in 1:2 ratio. The addition of (2-hydroxypropyl)-β-cyclodextrin to the background electrolyte was found to improve analysis performance. A Doehlert design, for the factors cyclodextrin concentration and voltage, was carried out and Derringer desirability function led to the identification of 18 mM and 29 kV as the optimal values. Applying the optimum conditions, separation of all the compounds, including the enantiomers of impurity 1, was obtained in less than 12 min. The method was validated according to ICH guidelines for drug assay and determination of impurities and was applied to oxybutynin tablet analysis.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 80, Issue 2, 15 December 2009, Pages 781–788