Fluorescence quantification of tetracycline in the presence of quenching matrix effect by means of a four-way model
چکیده انگلیسی

In this work, a four-way tensor is used to model the quenching effect in fluorescent measurements. By means of the analysis of excitation–emission matrices obtained in the determination of tetracycline in tea, which acts as quencher, it is shown as the impossibility to use a calibration, or an addition standard based on a three-way model. It is analysed the quencher multiplicative effect made on the tetracycline signal by means of an ANOVA. However, by arranging the experimental data in a four-way tensor, it is viable to perform a calibration based on the parallel factor analysis, PARAFAC, decomposition and a four-way partial least squares, 4-PLS, regression to quantify the tetracycline in the presence of the matrix quencher effect. 4-PLS calibration provides better results. In the range from 40 to 220 μg L−1 gives an average of relative errors in absolute value equal to 8.02% in prediction (3.40% in calibration). The repeatability as standard deviation in this range is 5.08 μg L−1 and the method is accurate, slope and intercept being statistically equal to 1 and 0, respectively when a regression calculated versus true concentration is performed. Moreover, it has a decision limit (CCα) of 13.87 μg L−1 for a probability of false positive, α, equal to 0.05 and a capability of detection (CCβ) of 26.63 μg L−1 (for probabilities of false positive, α, false negative, β, equals to 0.05).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 77, Issue 3, 15 January 2009, Pages 1129–1136
نویسندگان
, , ,